• Inventory
  • Products
  • Technical Information
  • Circuit Diagram
  • Data Sheet
Data Sheet
Home > Data Sheet > Q2004V31V
Q2004V31V

Q2004V31V

Model Q2004V31V
Description Thyristor Product Catalog
PDF file Total 224 pages (File size: 3M)
Chip Manufacturer TECCOR
AN1006
Application Notes
Model 577 Curve Tracer Procedure Notes
Because the curve tracer procedures in this application note are written for the Tektronix model 576 curve tracer, certain settings must
be adjusted when using model 577. Model 576 curve tracer has separate controls for polarity (AC,+,-) and mode (Norm, DC, Leakage),
whereas Model 577 has only a polarity control. The following table shows the guidelines for setting Collector Supply Polarity when
using model 577. (Figure AN1006.28)
Model 576
If using Leakage mode along with polarity setting of +(NPN) and -(PNP),
[vertical scale divided by 1,000],
If using DC mode along with either +(NPN) or -(PNP) polarity,
If using Norm mode along with either +(NPN) or -(PNP) polarity,
If using Norm mode with
AC
polarity,
Model 577
set
Collector Supply Polarity
to either
+DC
or
-DC,
depending on polarity setting
specified in the procedure. The vertical scale is read directly from the scale on the
control knob.
set
Collector Supply Polarity
to either
+DC
or
-DC
depending on polarity
specified.
set
Collector Supply Polarity
to either
+(NPN)
or
-(PNP)
per specified procedure.
set
Collector Supply Polarity
to
AC.
One difference between models 576 and 577 is the Step/Offset
Polarity setting. The polarity is inverted when the button is
depressed on the Model 576 curve tracer. The Model 577 is
opposite
the
Step/Offset Polarity is “inverted” when the button
is extended and “Normal” when the button is depressed. The
Step/Offset Polarity is used only when measuring I
GT
and V
GT
of
triacs and
Quadracs
in Quadrants l through lV.
Also, the Variable Collector Supply Voltage Range and Power
Dissipation controls have different scales than model 576. The
following table shows the guidelines for setting Power Dissipation
when using model 577.
Model 576
If power dissipation is 0.1 W,
If power dissipation is 0.5 W,
If power dissipation is 2.2 W,
If power dissipation is 10 W,
If power dissipation is 50 W,
If power dissipation is 220 W,
Model 577
set at 0.15 W.
set at 0.6 W.
set at 2.3 W.
set at 9 W.
set at 30 W.
set at 100 W.
Although the maximum power setting on model 576 curve tracer
is 220 W (compared to 100 W for model 577), the maximum col-
lector current available is approximately the same. This is due to
the minimum voltage range on model 577 curve tracer being
6.5 V compared to 15 V for model 576. The following table shows
the guidelines for adapting Collector Voltage Supply Range set-
tings for model 577 curve tracer procedures:
Model 576
If voltage range is 15 V,
Model 577
set at either 6.5 V or 25 V, depending on parameter
being tested. Set at 6.5 V when measuring V
TM
(to
allow maximum collector current) and set at 25 V
when measuring I
GT
and V
GT
.
set at 100 V.
set at 1600 V.
If voltage range is 75 V,
If voltage range is 1500 V,
http://www.teccor.com
+1 972-580-7777
AN1006 - 18
©2002 Teccor Electronics
Thyristor Product Catalog
Go Upload

* Only PDF files are allowed for upload

* Enter up to 200 characters.